Turbidity Analyzers-TB750G
2016-05-27 15:08:24
Description:
Since their sales began in 1959, Yokogawa's turbidity analyzer have been continuously developed and improved using various measurement principles suited for various applications. With its many achievements, Yokogawa has earned its customers' confidence. Developed based on years of experience and applications in process fields, the TB750G turbidity analyzer using right angle light scattering method provides highly reliable measurement and improved maintainability which improve upon what previous models could offer. A wide range of options are available to meet users' various needs.……
New Products
Honeywell 51307038-100 Power Control Module Assy
Honeywell 51307038-100 Power Control Module Assy. Honeywell Experion LX Series 8 C300 controller provides powerful and robust control for the Experion LX platform.
Honeywell 8937-HN2 Fibre Optic converter IO Module
Honeywell 8937-HN2 Multi-Mode, Fibre Optic converter IO Module. The Fiber Option Extender modules support the extension of the I/O link beyond the Series 8 cabinet with the S8 C300, enabling remote applications of the Series 8 I/O.
Honeywell 8939-HN Multi-Mode Fiber Optic Extender (IOTA)
Honeywell 8939-HN Multi-Mode Fiber Optic Extender IOTA. The Fiber Option Extender modules support the extension of the I/O link beyond the Series 8 cabinet with the S8 C300, enabling remote applications of the Series 8 I/O.
G.M. International X1-IS-AO-02-S I.S. SIL2 HART® Isolating Driver module
G.M. International X1-IS-AO-02-S I.S. SIL2 HART® Isolating Driver module. GMI Intrinsically Safe Isolators, Analog Output
Assortments
▶ Pressure Transmitter EJA
▶ Pressure Transmitters EJA-E
▶ Pressue Transmitter EJX
▶ Safety Instrumented System (SIS)
▶ Valve Positioners & Converters
▶ Field Mount Indicators
▶ Temperature Transmitters
▶ Magnetic Flowmeters
▶ Vortex Flowmeter
▶ Ultrasonic Flowmeters
▶ Coriolis Mass Flowmeters - ROTAMASS 3-Series
▶ Variable Area Flowmeters – Rotameters
▶ Device Smart Communicators
▶ PORTABLE AND HANDHELD INSTRUMENTS
▶ Distributed Control System (DCS)
▶ Process Analyzers
▶ Paperless Recorders
▶ Strip Chart Recorders
▶ Data Acquisition Equipment
▶ UT Series Temperature Controllers
▶ UP Series Program Controllers
▶ UM Series Digital Indicators with Alarms
▶ UT100 Series Temperature Controllers
▶ UD Series Manual Setters
▶ Limit Controllers
▶ YS Series Single-loop Controllers
Description
TB750G
Since their sales began in 1959, Yokogawa's turbidity analyzer have been continuously developed and improved using various measurement principles suited for various applications. With its many achievements, Yokogawa has earned its customers' confidence.
Developed based on years of experience and applications in process fields, the TB750G turbidity analyzer using right angle light scattering method provides highly reliable measurement and improved maintainability which improve upon what previous models could offer. A wide range of options are available to meet users' various needs.
Highly reliable measurement with excellent linearity and repeatability.
Easy-to-clean cell.
Compact, lightweight converter and detector.
User configurable measuring range: 0-0.2 NTU to 0-100 NTU.
Measuring range switching (2 or 3 ranges).
Enhanced self-diagnostic function as standard.
Detector structure to remove sudden reading change caused by bubbles.
A wide range of measurement conditions.
2 analog outputs, 3 relay contact outputs, and 1 serial communication.
Application
Turbidity of finished water and water used in general processes
| Measuring method | Right angle scattered light method |
| Measuring range | 0.000 to 100.0 NTU |
| Output range | Configurable within the measuring range Minimum range: 0 to 0.2 NTU Maximum range: 0 to 100 NTU Minimum span: 20 % or more of upper limit of the range or 0.2 NTU, whichever is greater. |
| Analog output | 2 outputs: output 1: 4 to 20 mA DC, isolated output 2: 4 to 20 mA DC or 0 to 20 mA DC, selectable, isolated (Maximum load resistance 550 Ω) |
| Serial communication | 1 output, RS-422 or RS-232C, isolated |
| Contact output | 3 contacts |
| Contact input | 2 contacts |
| Calibration | Zero/span calibration |
| Ambient temperature | -5 to 50℃ (Sample and tap water may need protection against freezing) |
| Ambient humidity | 5 to 95 % RH (non-condensing) |
| Sample water conditions | Flow rate: 0.05 to 20 l/min Temperature: 0 to 50℃ Pressure: 500 kPa max. |
| Power supply | 100 to 240 V AC -15 %/+10 %, 50/60 Hz |
| Power consumption | Converter + Detector: 50 VA max. |
| Repeatability | ±1 % of reading or ±0.002 NTU, whichever is greater |
| Linearity | ±2 % of reading or ±0.01 NTU, whichever is greater |
| Response time | Within 2 minutes (90 % response, sample water flow rate 3 l/min) |