Description:

The FC400G non-reagent free available chlorine analyzer adopts the polarographic method, using a rotating electrode to continuously measure online the concentration of free available chlorine. The FC400G is a microprocessor-based, intelligent free chlorine analyzer whose development was based on Yokogawa's long experience with residual chlorine analysis. The FC400G is intended for measurements in the water distribution network, at pumping stations, water delivery points, schools, and public facilities.……

Description

FC400G

The FC400G non-reagent free available chlorine analyzer adopts the polarographic method, using a rotating electrode to continuously measure online the concentration of free available chlorine.

The FC400G is a microprocessor-based, intelligent free chlorine analyzer whose development was based on Yokogawa's long experience with residual chlorine analysis.

The FC400G is intended for measurements in the water distribution network, at pumping stations, water delivery points, schools, and public facilities.

Features

  • One-touch simple calibration.

  • Output range can be arbitrarily set in the range of from 0-1 mg/l to 0-3 mg/l in the field.

  • Line segment output is available for output signals.

  • Easy confirmation of inter-electrode applied voltage/current characteristic (plateau characteristic). This allows understanding of the detailed sensor characteristics in the application.

  • Rotating electrode/ceramic bead cleaning has been improved to realize further effective and continuous automatic cleaning.

  • Easy-to-maintain (-clean) cell construction.

Specifications

Measured ObjectFree chlorine in tap water
Measuring SystemPolarographic method using rotating electrode
Measuring Range0 to 3 mg/l
Output RangeCan switch between arbitrary ranges (each between 0 to 1 mg/l and 0 to 3 mg/l)
Output Signal4 to 20mA DC (load resistance: 550 Ω or less) or 1 to 5V DC (output resistance : 300 Ω or less)
Sample ConditionsTemperature: 0 to 50℃
pH: 6.5 to 7.5
Flow Rate: 0.1 to 2.5 l/min
Pressure: 1 to 150 kPa
          100 to 740 kPa (when a dedicated sampling unit is provided)
Conductivity: 100 to 300 μS/cm
SS (Suspended Solid): 10 mg/l or less
ElectrodeIndicating electrode: Rotating gold electrode
Reference electrode: Silver electrode (with Pt1000 RTD installed)
Electrode CleaningCeramic bead cleaning
Ambient Temperature-30 to 70℃
InstallationIndoor (Outdoor use requires protection against rain.) (Direct sunlight must be avoided.)
Power Supply100/110/200/220 V AC ±10 %, 50/60 Hz
WeightApprox. 6 kg

Performance

% display shows value relative to the upper limit of a range.

Linearity±5 % of full scale
Repeatability2 % of full scale
StabilityZero drift; ±1 % of full scale/month or less
Span drift; -10 % of full scale/month or less
Response timeApprox. 2 min (90 % response time)