DIGITAL INSULATION TESTER MY40
2016-05-24 10:45:30
Description:
Digital model with 4 voltage/resistance ratings Multifunction Insulation resistance, AC voltage and conductor resistance measurement Insulation test mode: Comparator, memory, auto-hold and discharge functions All test modes: Live-line alarm (excluding AC voltage measurement), battery check and automatic power-off Easy-to-view, fluctuation-free display Double-action safety mechanism Weight including main unit and batteries only, excluding accessories……
INSULATION TESTER ACCESSORIES
Model Description
98007 Earth and Line Probe for 2406
B9646CA Test Lead Storage Bag (3213)
B9600HA Case for 3213
321803 Lead for Terminal Guard (2406E; 3213A)
B9600NZ Probe Tip Pointed #5
B9600NX Probe Tip Extension #3
B9600GN General Purpose Probe #1
B9600NV Probe Tip Storage Bag
98050 Line Probe and Earth Probe (3213A)
98002 My-Earth Probe
B9075MU Hard Case with Probe Case (2406D & E)
99005 My-Shoulder Strap
93013 My-Protection Cover
98001 My-Line Probe
New Products
FXT1-P37 Right Angle Milling Head
FXT1-P37 Right Angle Milling Head - HKXYTECH
FXT1-P24 Right Angle Milling Head
FXT1-P24 Right Angle Milling Head - HKXYTECH
FXT1-P18 Right Angle Milling Head
FXT1-P18 Right Angle Milling Head - HKXYTECH
FXT1-AC Fully Automatic Right Angle Milling Heads
FXT1-AC Fully Automatic Right Angle Milling Heads - HKXYTECH
Assortments
▶ Pressure Transmitter EJA
▶ Pressure Transmitters EJA-E
▶ Pressue Transmitter EJX
▶ Safety Instrumented System (SIS)
▶ Valve Positioners & Converters
▶ Field Mount Indicators
▶ Temperature Transmitters
▶ Magnetic Flowmeters
▶ Vortex Flowmeter
▶ Ultrasonic Flowmeters
▶ Coriolis Mass Flowmeters - ROTAMASS 3-Series
▶ Variable Area Flowmeters – Rotameters
▶ Device Smart Communicators
▶ PORTABLE AND HANDHELD INSTRUMENTS
▶ Distributed Control System (DCS)
▶ Process Analyzers
▶ Paperless Recorders
▶ Strip Chart Recorders
▶ Data Acquisition Equipment
▶ UT Series Temperature Controllers
▶ UP Series Program Controllers
▶ UM Series Digital Indicators with Alarms
▶ UT100 Series Temperature Controllers
▶ UD Series Manual Setters
▶ Limit Controllers
▶ YS Series Single-loop Controllers
Description
Specifications - MY40 Digital Insulation Tester
Width | 125 mm |
---|---|
Height | 103 mm |
Depth | 53 mm |
Display | 3 1/2-digit LCD; 4000 count; backlight-illuminated; logarithmic bar graph; extension bar graph-no fluctuations, as the display shows the digits of a reading in the order in which each digit settles. |
Comparator Function | The MY40 alerts you by turning on the LOW symbol and sounding the buzzer if the measured value is smaller than the reference value. You can allocate as many as three user-defined reference values to each rating. The factory-set defaults are 0.1 MΩ, 0.2 MΩ and 0.4 MΩ. |
Memory Function | For each rating, you can save as many as 20 measurements at desired memory address numbers. |
Automatic Discharge Function | The MY40 automatically begins discharge when you turn off the MEAS switch. You can monitor the state of discharge by checking the bar graph and make sure discharge is complete by checking that the segment bar disappear. |
High-Voltage Indicators | The high-voltage symbol and LED lamp come on to alert you when the MY40 is in insulation testing mode or if any voltage remains to be discharged. |
Live-Line Alarm | If you apply an AC voltage of approximately 40 V or higher across the input terminals, the MY40 alerts you by blinking the LED lamp and sounding the buzzer. |
Overrange Input Alarm | If the voltage being measured exceeds 600 V during AC voltage measurement, the MY40 alerts you by flashing the Maximum Value indicator and sounding the buzzer. |
Auto-Hold Function | The tester retains the measured resistance for approximately 5 seconds after the MEAS switch is turned off. |
Batteries | Four AA (R6P) batteries |
Weight | 420 g |
Ambient Temperature/Humidity Ranges | 23 ±5 /45-75% RH |
Standard Accessories | Batteries (Qty - 4) Earth Probe (98002) (Qty - 1) Line Probe (98001) (Qty - 1) Protection Cover (93013) (Qty - 1) Shoulder Strap (99005) (Qty - 1) User's manual (Qty - 1) |
Tolerance Mentioned Conditions
Max. Deviation from Zero Scale Value | 6 digits |
---|---|
Indication of mark on bar graph | Approx. 4000 MΩ min. (500 V/1000 V) Approx. 400 MΩ min. (125 V/250 V) |
No-Load Voltage | 130% max. of rated voltage |
Rated Measuring Current | 1 mA (0 to 20%) when in first effective measuring range |
Max. Short-circuit Current | 2 mA |
Testing Performance
Model | Rating | Range Option | Resolution | Measuring Range | Tolerance | Lower Limit of measured Ω | Rated Current | Central Scale Value |
---|---|---|---|---|---|---|---|---|
MY40-01 | 125V/200MΩ | 0.4000 | 0.1kΩ | 0-0.0199MΩ | ± (5%of rdg+6dgt) | 0.125MΩ | 1mA | 5MΩ |
4.000 | 1kΩ | 0.0200-20.00MΩ* | ± (2%of rdg+6dgt) | |||||
40.00 | 10kΩ | 20.01-200.0MΩ | ± 5%of rdg | |||||
200.0 | 100kΩ | |||||||
250V/200MΩ | 0.4000 | 0.1kΩ | 0-0.0499MΩ | ±(5%of rdg+6dgt) | 0.25MΩ | 1mA | 5MΩ | |
4.000 | 1kΩ | 0.0500-20.00MΩ* | ± (2%of rdg+6dgt) | |||||
40.00 | 10kΩ | 20.01-200.0MΩ | ±5%of rdg | |||||
200.0 | 100kΩ | |||||||
500V/2000MΩ | 4.000 | 1kΩ | 0-0.999MΩ | ± (5%of rdg+6dgt) | 0.5MΩ | 1mA | 50MΩ | |
40.00 | 10kΩ | 1.000-500MΩ* | ± (2%of rdg+6dgt) | |||||
400.0 | 100kΩ | 501-2000MΩ | ± 5%of rdg | |||||
2000 | 1MΩ | |||||||
1000V/2000MΩ | 4.000 | 1kΩ | 0-1.999MΩ | ± (5%of rdg+6dgt) | 2MΩ | 0.5mA | 50MΩ | |
40.00 | 10kΩ | 2.000-1000MΩ* | ± (2%of rdg+6dgt) | |||||
400.0 | 100kΩ | 1001-2000MΩ | ± 5%of rdg | |||||
2000 | 1MΩ |