Turbidity Analyzers-TB750G-hkxytech
  • Turbidity Analyzers-TB750G

Turbidity Analyzers-TB750G

2016-05-27 15:08:24

Description:

Since their sales began in 1959, Yokogawa's turbidity analyzer have been continuously developed and improved using various measurement principles suited for various applications. With its many achievements, Yokogawa has earned its customers' confidence. Developed based on years of experience and applications in process fields, the TB750G turbidity analyzer using right angle light scattering method provides highly reliable measurement and improved maintainability which improve upon what previous models could offer. A wide range of options are available to meet users' various needs.……

Notice:

New Products

Honeywell Analytics SPSTAXH1SS Sensepoint Flammable and Toxic Gas Detector Explosion-Proof Transmitters

Cost-effective, fixed point gas detector for the protection of personnel and plant from flammable, toxic and Oxygen gas hazards

Honeywell Analytics Sensepoint XCD RTD Remote toxic and oxygen gas detector for industrial applications

3 wire, 4-20mA and RS485 MODBUS output fixed point detector with in-built alarm and fault relays for the protection of personnel and plant from toxic and oxygen hazards.

Honeywell Analytics Sensepoint XCD RFD Remote Flammable Gas Monitor Explosion-Proof Transmitters

3-wire, 4-20mA, transmitter with optional Modbus for use with remotely mounted flammable gas sensors 705 and 705HT

Assortments

Description

TB750G

Since their sales began in 1959, Yokogawa's turbidity analyzer have been continuously developed and improved using various measurement principles suited for various applications. With its many achievements, Yokogawa has earned its customers' confidence.

Developed based on years of experience and applications in process fields, the TB750G turbidity analyzer using right angle light scattering method provides highly reliable measurement and improved maintainability which improve upon what previous models could offer. A wide range of options are available to meet users' various needs.

  • Highly reliable measurement with excellent linearity and repeatability.

  • Easy-to-clean cell.

  • Compact, lightweight converter and detector.

  • User configurable measuring range: 0-0.2 NTU to 0-100 NTU.

  • Measuring range switching (2 or 3 ranges).

  • Enhanced self-diagnostic function as standard.

  • Detector structure to remove sudden reading change caused by bubbles.

  • A wide range of measurement conditions.

  • 2 analog outputs, 3 relay contact outputs, and 1 serial communication.

Application

  • Turbidity of finished water and water used in general processes

Measuring methodRight angle scattered light method
Measuring range0.000 to 100.0 NTU
Output rangeConfigurable within the measuring range
Minimum range: 0 to 0.2 NTU
Maximum range: 0 to 100 NTU
Minimum span: 20 % or more of upper limit of the range or 0.2 NTU, whichever is greater.
Analog output2 outputs:
output 1: 4 to 20 mA DC, isolated
output 2: 4 to 20 mA DC or 0 to 20 mA DC, selectable, isolated (Maximum load resistance 550 Ω)
Serial communication1 output, RS-422 or RS-232C, isolated
Contact output3 contacts
Contact input2 contacts
CalibrationZero/span calibration
Ambient temperature-5 to 50℃ (Sample and tap water may need protection against freezing)
Ambient humidity5 to 95 % RH (non-condensing)
Sample water conditionsFlow rate: 0.05 to 20 l/min
Temperature: 0 to 50℃
Pressure: 500 kPa max.
Power supply100 to 240 V AC -15 %/+10 %, 50/60 Hz
Power consumptionConverter + Detector: 50 VA max.
Repeatability±1 % of reading or ±0.002 NTU, whichever is greater
Linearity±2 % of reading or ±0.01 NTU, whichever is greater
Response timeWithin 2 minutes (90 % response, sample water flow rate 3 l/min)