  • Yokogawa Separate type Zirconia Oxygen Analyzer, Converter

Yokogawa Separate type Zirconia Oxygen Analyzer, Converter

2016-05-25 17:37:57

Description:

Models: ZR22G, ZR22S, ZR402G, ZR22P, ZR22R, ZR22V, AC1, MC1, M1233SR, IAC-24, WZ-H, M1234SE-A, M1200DB-06 ……

Yokogawa's zirconia oxygen analyzers are used for combustion monitoring and control in a wide variety of applications and help industries to make considerable achievements in energy saving. The applications range from energy-consuming industries, such as iron and steel, electric power, oil and petrochemical, ceramics, pulp and paper, food, and textiles, to various combustion facilities, such as incinerators and small- and medium sized boilers. Combustion monitoring and control using the analyzer also contribute to a lowering of CO2, SOx, and NOx emissions by allowing for more complete combustion, thus preventing global warming and air pollution.

Description

Model Separate type Zirconia Oxygen Analyzer, Converter

The ZR402G converter is equipped with an LCD touch screen which has various setting displays, a calibration display, oxygen concentration trend display, with easier operation and improvement of display functions. The converter is equipped with various standard functions such as measurement and calculation as well as maintenance functions including self-test. The ZR22G separate-type detector uses a high-reliability zirconia sensor, and its heater assembly can be replaced on site. The detector is mounted, for example, on the wall of a flue and can measure the gases directly. The ZR22S separate-type detector for explosion-proof applications is also available.


  • The separate type converter ZR402G incorporates a LCD touchscreen for ease of operation.

  • The built-in heater assembly of the probe can be replaced on site, reducing maintenance costs (ZR22G only).

  • The probe uses a long-life, high-reliability, field replacable Zirconia sensor.

  • Remote maintenance using digital communications (HART) reduces maintenance costs.

Note: HART is a registered trademark of HART Communication Foundation.

Measurement objectOxygen concentration in combustion exhaust gas and mixed gas
Measurement systemZirconium oxide sensor
Measurement range0.01 to 100 vol% O2
Output signal4 to 20 mA DC, two points (ZR402G) (maximum load resistance 550Ω)
Setting rangeAny setting in the range of 0 - 5 vol%O2 to 0 - 100 vol%O2 (in 1 vol% O2), or partial range
Digital communication (HART)250 to 550Ω, depending on number of field devices connected to the loop (multi-drop mode)
Sample gas pressure-5 to +250 kPa (Pressure compensation option necessary over +5 kPa)
Sample gas temperature0 to 700℃ (Probe only)
700 to 1871℃ (with 0.15m probe with JIS 5K 32 FF flange and ZR22P High Temperature Probe Adapter)
Probe length0.15, 0.4, 0.7, 1.0, 1.5, 2.0, 2.5, 3.0, 3.6, 4.2, 4.8, 5.4 m
Contact outputFour points (one is fail-safe, normally open)
Contact inputTwo points
CalibrationMethod; zero/span calibration
Calibration mode; automatic, semi-automatic and manual (with an LCD touchscreen)
Construction of detector (ZR22G)Heater and thermocouple replaceable construction. 
Non-explosion-proof. Equivalent to NEMA4X/IP66
Construction of converter (ZR402G)Non-explosion-proof, outdoor installation, equivalent to NEMA4X/IP66
Ambient TemperatureProbe (ZR22G): -20 to 150℃
Converter (ZR402G): -20 to 55℃)
Reference Gas System

Instrument Air (preferred): 50 kPa (200 kPa with check valve) + the pressure inside the furnace; Approx. 1 Nl/min consumption

Natural convection

Power supply85 to 264 V AC; 45 to 66 Hz

Measurement objectOxygen concentration in combustion exhaust gas and mixed gas
Measurement systemZirconium oxide sensor
Explosion-proof ApprovalZR22S-A (ATEX); EEx d IIB + H2, 
ZR22S-B (FM); Class I, Division1, Groups B, C and D
ZR22S-C (CSA); Class I, Division1, Groups B, C and D
ZR22S-D (IECEx); Ex d IIB + H2, T2
Measurement range0.01 to 100 vol% O2
Output signal4 to 20 mA DC, two points (ZR402G) (maximum load resistance 550Ω)
Setting rangeAny setting in the range of 0 - 5 vol%O2 to 0 - 100 vol%O2 (in 1 vol% O2), or partial range
Digital communication (HART)250 to 550Ω, depending on number of field devices connected to the loop (multi-drop mode)
Sample gas pressure±5kPa
For 0.15m probe, -0.5 to +5 kPa.
No pressure fluctuation in the furnace should be allowed
Sample gas temperature0 to 700℃
700 to 1871℃ (with 0.15m probe with JIS 5K 32 FF flange and ZR22P High Temperature Probe Adapter (certain options only))
Probe length0.15, 0.4, 0.7, 1.0, 1.5, 2.0 m
Contact outputFour points (one is fail-safe, normally open)
Contact inputTwo points
CalibrationMethod; zero/span calibration
Calibration mode; automatic, semi-automatic and manual (with an LCD touchscreen)
Construction of detector (ZR22S)Heater and thermocouple replaceable construction, equivalent to NEMA4X/IP66
Construction of converter (ZR402G)Non-explosion-proof, outdoor installation, equivalent to NEMA4X/IP66
Ambient Temperature-20 to +60℃ (-20 to +150℃ on the terminal box surface) (ZR22S)
-20 to +55℃ (ZR402G)
Reference Gas SystemInstrument Air: 50 kPa + the pressure inside the furnace; Approx. 1Nl/min consumption
Wiring ConnectionATEX; M20 by 1.5 mm or 1/2 NPT select one type (4 pieces)
FM; 1/2 NPT (4 pieces)
CSA; 1/2 NPT (4 pieces)
IECEx; M20 by 1.5 mm or 1/2 NPT select one type (4 pieces)
Power supply85 to 264 V AC; 45 to 66 Hz

Repeatability±0.5% Maximum value of set range
Linearity±1% Maximum value of set range
Drift±2% Maximum value of set range/month
Response time (T90)90% response within 5 sec (60 sec for explosion-proof ZR22S)

ZR402G Model codes.png


System Components

ZR22S Separate type Explosion-proof Zirconia Oxygen Analyzers, Detector

ZR402G Separate type General use Zirconia Oxygen Analyzer, Converter

ZR202S Integrated type Explosion-proof Zirconia Oxygen Analyzers

ZO21P High Temperature Probe Adapter for separate type Zirconia Oxygen Analyzer

E7046EC, E7046EN Ejector Assembly for High Temperature

ZO21R Probe Protector for Zirconia Oxygen Analyzers

ZO21S Standard Gas Unit (*2)

ZA8F Flow Setting Unit for manual calibration

ZR40H Automatic Calibration Unit for Separate type Analyzer

Automatic Calibration Unit for Integrated type Oxygen Analyzer (*1)

L9852CB, G7016XH Stop Valve for Calibration gas line

K9292DN, K9292DS Check Valve for Calibration gas line

G7003XF/K9473XK, G7004XF/K9473XG Air Set

G7013XF, G7014XF Pressure Regulator for Gas Cylinder

ZR22A, ZR202A Heater Assembly for Spare Parts

Models:


ZR402G-T-E-E-A,ZR402G-P-E-E-A,ZR402G-G-E-E-A,ZR402G-M-E-E-A

ZR402G-T-E-E-A/SCT,ZR402G-P-E-E-A/SCT,ZR402G-G-E-E-A/SCT,ZR402G-M-E-E-A/SCT

ZR402G-T-E-E-A/PT,ZR402G-P-E-E-A/PT,ZR402G-G-E-E-A/PT,ZR402G-M-E-E-A/PT

ZR402G-T-E-E-A/H,ZR402G-P-E-E-A/H,ZR402G-G-E-E-A/H,ZR402G-M-E-E-A/H

ZR402G.jpgPA091193.jpg