Description:
Metrix HI 803 Deluxe Hardy Shaker HI 803-120V,The Hardy Shaker ensures that the Seismic Sensors and Transmitters that you have installed are working properly and giving you the results you need. Just like the pressure and flow instruments in your facility need calibration checks every year, so do your Seismic Sensors and Transmitters. This system puts the calibration check of those Seismic Sensors and Transmitters within your control, and within your budget. ……
Metrix HI 803 Deluxe Hardy Shaker HI 803-120V,brand new, origin in USA;
New Products
FXT1-P37 Right Angle Milling Head
FXT1-P37 Right Angle Milling Head - HKXYTECH
FXT1-P24 Right Angle Milling Head
FXT1-P24 Right Angle Milling Head - HKXYTECH
FXT1-P18 Right Angle Milling Head
FXT1-P18 Right Angle Milling Head - HKXYTECH
FXT1-AC Fully Automatic Right Angle Milling Heads
FXT1-AC Fully Automatic Right Angle Milling Heads - HKXYTECH
Assortments
Description
Metrix HI 803 Deluxe Hardy Shaker HI 803-120V
Metrix HI 803 Deluxe Hardy Shaker HI 803-120V
The Hardy Shaker ensures that the Seismic Sensors and Transmitters that you have installed are working properly and giving you the results you need. Just like the pressure and flow instruments in your facility need calibration checks every year, so do your Seismic Sensors and Transmitters. This system puts the calibration check of those Seismic Sensors and Transmitters within your control, and within your budget.
The most cost effective maintenance strategy is a predictive and proactive strategy, versus a preventative maintenance or run-to-failure strategy. The entire premise of a predictive and proactive maintenance program is based on the reliability and integrity of the transducers providing the information in a timely way. If the Seismic transducers are not working properly, you will not receive the value you are expecting in order to produce the return on investment that you have already made. The Hardy Shaker is a necessary tool to ensure your proactive maintenance strategy is effective.
Overview
Model:HI803
This top-of-the-line Hardy shaker includes more operational functions. In manual mode, the vibration test level and test frequency can be set from 10 Hz to 10,000 Hz. In sensitivity mode, the HI 803 not only vibrates the transducer being tested, but also measures its output and calculates its output sensitivity. A built-in display reads the output sensitivity, which can be stored in memory. Complete testing sequences and data can be stored in memory for each tested sensor, and a built-in printer logs data on the spot or stores it for later. An RS232 port and data transfer program are standard, and ASCI II test data files are easily transferred to a computer. For testing of heavy transducers or at high frequencies, the HI 803 automatically corrects the sensitivity calculation for the effects of transducer mass loading. It also measures and displays transducer phase, which is important in balancing applications. A built-in self test program checks for accuracy of all calculations.
Product Aprox. Weight: 9.5 kg (21 lbs)
HI 803/813 portable vibration shakers are designed to be used in the field or shop to verify the performance of vibration transducers, connectors, cabling, instruments or permanently installed condition monitoring systems. Both models are battery-powered, portable and feature a built-in reference accelerometer and indicator with calibration traceable to the National Institute of Standards and Technology (NIST).
These shakers can easily check velocity transducers, accelerometers and non-contacting Eddy Probe displacement measurement systems.
Simply attach the transducer to the shaker head (displacement probes require a mounting fixture and target), set the shaker to a known vibration reference level, and verify readings. The reference test level can be set in acceleration, velocity, or displacement, English or Metric.
FEATURES AND BENEFITS
• Actual precision vibration testers
• True and accurate testing of vibration sensors
• NIST traceable calibration reference
• High-level vibration accuracy
• Variable frequency and adjustable vibration amplitudes
• Rapid set-up for all vibration conditions
• Battery operated
• Ability to “take the calibration lab to the field”
APPLICATIONS
• Loop checks of vibration monitoring systems
• Verification and testing of velocity transducers,
accelerometers, eddy current probes, etc.
• Troubleshooting of issues in vibration monitoring loops
• Calibration of portable vibration analyzers
• Confirmation of vibration alarms
SPECIFICATIONS
HI 803
Battery Type Sealed, rechargeable
Charger Internal: Specify 120 or 240 Vac
Frequency Range Acceleration Readout (+ 1 digit)
10 to 10,000 Hz
10 to 2,000 Hz, +.3 dB
2,000 to 10,000 Hz,
+1 dB
Velocity Readout 10 to 500 Hz, +.3 dB
Displacement Readout: 25 to 200 Hz, +.3 dB
Maximum Output: 10g at 100 Hz, 100 gram load (3g above 1,000 Hz) 1g at 100 Hz, 750 gram load
Calibration:NIST traceable standard
Waveform Distortion: <5% THD, 30 to 2,000 Hz
Maximum Unsupported Load: 750 grams, vertical or 500 grams, horizontal
Physical Size & Weight: 11”W x 7”D x 10”H, 21 lbs.
Sensor Mounting Thread: 1/4-28 thd hole, 1/4” deep
English & Metric Readout: Yes
Indicates Test Sensor Output: Yes, normalized sensitivity mass loading compensation, phase
Internal Memory: Yes, stores test data
Built-in Printer: Yes, thermal type,prints current or stored test data
RS-232 Output: Yes, stored data transfer
ACCESSORIES
PART NO. DESCRIPTION
6006-0008 Power Cord, AC
2824-0135 Screw, 1/4-28X
0228-0071-01 Mounting Fixture, Accelerometer
5407-0001 Screwdriver, Flat Blade
0202-0072-01 Communication Software
0215-0183-01 Charge Cable
2112-0007 Cable Connector Kit, HI 803 input from TUT
0215-0129-01 Cable, Portable Shaker to 2-pin HI 113 Accel
6006-0008 Power Cord, AC
0228-0070-01 Stud, 1/4-28 to 1/4-28 UNF
0228-0070-03 Bushing, 1/4-28 to 2-56 UNC
0228-0070-04 Bushing, 1/4-28 to 6-32 UNC
0228-0070-05 Bushing, 1/4-28 to 10-32 UNC
0228-0071-02 Velocity/Universal Transducer Adapter, w/screw
0228-0072-01 Wrench, Shaker Head
NON STANDARD ACCESSORIES
HI 224 Eddy Current Probe Test Kit
0205-0049 Mounting Spacer
0228-0136-01 Mounting Plate, Magnetic 2” Dia, 1/4”thick
1901-0057 Paper, Thermal Printer (Seiko p/n 201-211-25C) HI 803 Calibration Kit
0228-0141-01 Stud, 1/4-28 to M8 (Metric)
0228-0137-01 Adapter stud. 1/4-28 UNF to 3/8-24 (HI5701VT)
0228-0075-01 4140 Steel Target (incl w/ 223/224 kits)
0228-0071-02 Velocity/Universal Transducer Adapter, w/screw
MODEL DESCRIPTION
HI 803-120V:
Portable Shaker, variable frequency and amplitude control, manual and sensitivity modes (sensitivity mode tests the transducer and provides a mV/unit output to compare with the sensor manufacturer’s specifications), built-in self-test program, built-in printer with paper, data storage, RS-232 port and data transfer software, 120V, standard accessories.
HI 803-240V
Same as the HI 803-120V, except the voltage requirement is 240 volts.
HI 813-CE